This article describes gas cross sensitivity, how it can impact your gas sensors, and what gas sensors can exhibit it. The article applies to G7c, G7x, and G7 EXO.
Cross sensitivity is a phenomenon that occurs when a gas other than the gas being monitored causes the gas sensor to show a reading, even when the target gas is not present. An example of this would be an H₂S sensor responding to H₂ in the area.
To alleviate cross sensitivity, Blackline:
During calibration and bump testing a cross sensitive sensor, you will see readings on the sensor you are not calibrating or bump testing because the sensor has a response to the gas being applied during the calibration or bump testing.
To reduce these cross sensitive responses, Blackline strongly recommends following the prescribed order of the gases applied to the device when doing either Dock or manual calibration or bump testing. For most sensor/gas combinations, the cross sensitive response can be removed by exposing the device to fresh air. Please consult with Customer Care for recommended fresh air exposure times.
Almost all gases cause some cross sensitivity, but our devices are designed to filter out many of the most commonly occurring ones.
The following sensors can have cross sensitivity.
Let us know if you have any questions — don’t hesitate to reach out to our Customer Care team.