Although the G7 EXO pump inlet filters are an optional accessory, Blackline recommends always installing the inlet filters when using G7 EXO Pump to maintain the pump in good condition.
NOTE: EXO pump does not support ClO2 or Cl2.
Using G7 EXO Pump without attaching inlet filters directly onto the inlets can allow water to be pulled into the pump.
WARNING: Inlet filters are for use with G7 EXO Pump model only. Do not use inlet filters with G7 EXO Diffusion.
For this procedure, you will need:
WARNING: Do NOT turn EXO pump on until you have properly fitted the inlet filters to the open the pump inlet spring valves. If the inlet filters are not properly fitted and EXO is turned on, you may see a “Pump Blocked” error.
NOTE: Hoses can be connected to multiple inlets at the same time, but EXO will only run one pump inlet at a time.
In humid environments or when moving from hot to cold temperatures, condensation may form in the hose. This condensation can freeze and cause a “Pump low flow” error on G7 EXO Pump. To resolve, try the following: